Oxygen Analyzer

LC-860

High Precision Oxygen Analyzer

High Precision and High Performance Measurement

연구・실험용 고성능 산소농도계   LC-860

 고성능 지르코니아식 산소농도계 LC-860은 당사의 오랜 시간에 걸친 경험과 기술을 바탕으로 개발되어 ppb 부터 100% 고농도까지 1대로 측정할 수 있는 고성능 산소농도계입니다.

본 산소농도계는 장시간에 걸쳐 안정적이고 정밀한 측정을 실현할 수 있는 당사의 독자적인 지르코니아 센서를 탑재하여 고성능을 실현하였습니다.

또한 기능별 키 입력으로 조작이 간단하고 고성능으로 설계되어 연구・실험용 및 온라인 장비에 적합한 산소농도계입니다.

Features

항 목 내 용
Measurement Range ppb to 100% O2 level
Easy Operation Easy to operate by functional key
Purge Function More precise measurement is available by more stable reference gas. (with purge function only)
High Reliability Sensor Deterioration-free in the sensor is also available (Optional)
Auto Range It enables to measure from high to low concentration.
Measurement Alarm Selectable for Normal Open or Closed.
Alarm output for upper / lower limit is selectable.
Rapid Response Rapid response is realized by bypass out and flow.
External Driving for Pump Remote ON/OFF is available.
Compliant standard Compliant with CE marking, EN61010-1, EN61326-1, and RoHS

 

Specification

항 목 내 용
Type Desktop or Panel Mounted
Display LED Digital 6 Digits
Meas. Range 1ppb~100vol% O2
Auto Range or Fixed Ranges
Analog Output Range 0-10/100/1,000ppm/100%
Repeatability ±0.5% F.S (More than 1% range)
±1.0% F.S (Less than 1% range)
Gas Response Within 10 Seconds (90% Response)
Sampling Method Continuous Suction by Built-in Pump
Sample Gas Flow 1,000 ~ 2,000 ml/min
Sensor Flow 200±50ml/min
Gas Connection Port PT 1/8”(F) (Inlet, Bypass out)
PT 1/4”(F) (Outlet)
Reference Gas Atmospheric Gas
Warm-up Time Within 20 min
Dimension 288(W)×144(H)×400(D)  Unit : mm
Weight Approx. 10kg
Analog Output DC 4~20mA and DC 0~10V
Transmission Output RS-232C Compliant , 9-Pin D-sub Male
Alarm Output High / Low alarm
Fail alarm (Instrument / Concentration Error)
Ready (Warm-up signal)
(No-voltage Contact / Contact capacity : 24VDC/AC or less, 0.1A or less)
Range Marker Output 3 Points Output
(No-voltage Contact / Contact capacity : 24VDC/AC or less, 0.1A or less)
Power Source AC220V±10V 50/60Hz, 200VA
Sample Gas Conditions Maximum permissible pressure: 30kPa(0.03MPa) Max
Humidity : Dew point at or below ambient temperature
Temperature : 60°C or less
Installation Conditions Location : Indoors, Non-Explosion Area
Ambient Temperature : 0~40℃
Ambient Humidity : 45~85%RH, No Condensation

 